Possible Cause |
Solution |
Comments |
Change in detector response |
Check gas flows, temperatures, and settings.Check background level or noise |
All peaks may not be equally affected.May be caused by system contamination and not the detector |
Co-elution |
Change column temperature or stationary phase |
Decrease column temperature and check for appearance of a peak shoulder or tail |
Change in the purge activation time |
Check the purge activation time
|
For splitless injection |
Change in the injection volume |
Check injection technique |
Injection volumes are not linear |
Change in sample concentration |
Check and verify sample concentration |
Sample degradation, variances in sample temperature or ph |
Leak in the syringe |
Use a different syringe |
Sample leaks passed the plunger or around the needle, Leaks are often not readily visible |
Column contamination |
Trim the column Solvent rinse the column |
Remove about half to 1 meter from the front of the column. Only bonded and crosslinked phases |
Column activity |
Irreversible |
Only affects active compounds |